Repeatability of gray value-based bone density measurements in cone beam computed tomography (CBCT) images under different acquisition protocols

不同采集方案下锥形束计算机断层扫描(CBCT)图像中基于灰度值的骨密度测量的重复性

阅读:1

Abstract

BACKGROUND: Accurate and repeatable assessment of bone density is a critical factor in dental implant planning, as it influences decisions related to implant size, site preparation, and primary stability, ultimately affecting treatment outcomes and long-term osseointegration. The variability in Cone-Beam Computed Tomography (CBCT) gray values across different acquisition protocols can compromise measurement repeatability, challenging the reliability of preoperative bone assessments. This study aimed to assess the repeatability of a gray value-based bone density measurement tool along three-thirds of a virtual implant in CBCT under different acquisition protocols. METHODS: Three pork ribs (5 × 3 × 2 cm) were scanned three times each using a VistaVoxS 3D CBCT scanner under four acquisition protocols, varying field-of-view (5 × 5 and 13 × 5), kVp (98, 94, 79), mA (11, 9, 6), and voxel size (0.08 and 0.2 mm), resulting in 144 scans. A virtual implant (10 × 5.2 mm) was centered on each rib image, and gray values along the thirds of the implant were measured using AIS 3DAPP 5.0 software. Repeatability was assessed with Friedman's test, Kendall's agreement coefficient, and the chi-square test. RESULTS: The mean CBCT gray values in the cervical region were significantly lower regardless of the protocol (p ≤ 0.5). Individual exposures differed but showed very good agreement (0.89-0.83). CONCLUSIONS: The bone density tool showed reliable gray value measurements across CBCT protocols, with field of view affecting repeatability more than voxel size. Significant gray value differences in the cervical region highlight the need for cautious interpretation in this area. Standardizing imaging settings is recommended to improve implant planning accuracy.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。