Does implant drill design influence the accuracy of dental implant placement using static computer-assisted implant surgery? An in vitro study

种植钻头设计是否会影响静态计算机辅助种植手术中种植体植入的精度?一项体外研究

阅读:1

Abstract

BACKGROUND: The purpose of this in vitro study was to compare the accuracy of implant placement in model surgeries according to the design of the drills (straight drills or step drills) used to finalize the implant bed during pilot-guided static computer-assisted implant surgery (sCAIS). METHODS: Model surgeries were carried out on resin models randomly assigned to three study groups. Virtual planning software (coDiagnostiX 10.6, Dental Wings, Montreal, Canada) was used to plan the implant positions. In Groups 1 and 2, pilot-guided sCAIS was performed. Straight drills were used in Group 1, and step drills were used in Group 2 to finalize the implant beds. In Group 3, fully guided sCAIS was performed using a universal fully guided kit (RealGUIDE Full Surgical Kit 3DIEMME, RealGUIDE, Cantù, Como, Italy). A total of 90 dental implants (Callus Pro, Callus Implant Solutions GmbH, Nuremberg, Germany) were placed (six implants per model, five models per study group). The primary outcome variables (angular deviation, coronal global deviation, and apical global deviation) were calculated for all implants based on the comparison of the preoperative surgical plan with the postoperative scans. RESULTS: Group 2 (coronal global deviation, 0.78 ± 0.29 mm; apical global deviation, 1.02 ± 0.56 mm) showed significantly lower values of both global deviation variables than Group 1 (coronal global deviation, 0.95 ± 0.20 mm; apical global deviation, 1.42 ± 0.49 mm). However, there was no significant difference in angular deviation between Groups 1 and 2 (7.56 ± 2.92° and 6.44 ± 2.84°). Group 3 produced significantly lower values of all three primary outcome variables (angular deviation, 2.36 ± 0.90°; coronal global deviation, 0.59 ± 0.28 mm; apical global deviation, 0.90 ± 0.29 mm) than Group 1 and significantly lower angular deviation and coronal global deviation values than Group 2. CONCLUSIONS: The design of the drills used to finalize implant osteotomies during pilot-guided sCAIS influences dental implant placement accuracy. Using step drills instead of straight drills for final osteotomies decreases deviation from the surgical plan. The fully guided approach performed better than the pilot-guided sCAIS.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。