A comprehensive study on the mechanical effects of implant-supported prostheses under multi-directional loading and different occlusal contact points

对种植体支持式义齿在多方向载荷和不同咬合接触点下的力学效应进行综合研究

阅读:1

Abstract

AIMS: To evaluate screw loosening and fracture load and angular deviation of a single implant-supported prosthesis under multi-directional loading condition at three different occlusal contact points. METHODS: A total of 40 metal crowns were cemented to external connection implants and were embedded vertically and obliquely. The occlusal surface of the crown was designed with three flat surfaces, contact a, b, and c, representing outer and inner 20-degree inclination for buccal and lingual cusps. The angular deviations of implant crown under static 50N of loading were measured. And screw removal torque was evaluated before and after 57,600 load cycles. Then, fracture load was measured for each specimen. Data analysis was performed using one-way analysis of variance test of significance followed by Tukey honest significant difference (HSD) test(p < 0.05). RESULTS: Angular deviation results showed statistical significance between all contact points in vertically embedded group compared to obliquely embedded group, which showed similar results between contact A and B compared to C. In the other hand, screw loosening evaluation did not show statistical significance among the tested groups. And for the fracture load evaluation the maximum values reached twice the yield values in all contact areas. CONCLUSIONS: Mechanical effects were different regarding to diverse loading direction and contact points. The results of this study suggest that the stress concentration might increase in unfavorable vector direction.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。