Fatigue Resistance of Cast-on Implant Abutment Fabricated with Three Different Alloys

三种不同合金铸造种植体基台的抗疲劳性能

阅读:1

Abstract

OBJECTIVES:  This study aimed to evaluate fatigue resistance of cast-on implant abutment using three alloys. MATERIALS AND METHODS:  Forty specimens of implant-supported crowns were prepared; Group 1 (TA) stock titanium abutments, Group 2 (GS) abutment cast with 40% gold alloy, Group 3 (GP) abutment cast with palladium alloy, and Group 4 (CN) abutment cast with nickel-chromium alloy. Specimens were cyclic loaded at 20 Hz, starting from 200 N (5,000 cycles), followed by stepwise loading of 400, 600, 800, 1,000, 1,200, 1,400, 1,600, and 1,800 N (30,000 cycles/step). Specimens were loaded until failure or reached 245,000 cycles. STATISTICAL ANALYSIS:  The withstand cycles were analyzed using one-way analysis of variance and Weibull survival analysis. Fracture surfaces were examined using scanning electron microscopy. RESULTS:  The results of withstand cycles were TA (189,883 ± 22,734), GS (195,028 ± 22,371), GP (187,662 ± 22,555), and CN (200,350 ± 30,851). The statistical analysis showed no significant difference between the groups (p = 0.673). CONCLUSION:  Although CN has higher Weibull characteristic strength which means greater durability, its lower Weibull modulus demonstrated less structural reliability. Consistent failures at implant fixture level were also found in CN group.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。