In vitro Intestinal Epithelial Wound-healing Assays Using Electric Cell-Substrate Impedance Sensing Instrument

利用细胞基质阻抗传感仪进行体外肠上皮伤口愈合试验

阅读:5
作者:Olivier Merlin-Zhang, Junsik Sung, Emilie Viennois

Abstract

Here, we describe an in vitro epithelial wound-healing assay using Electric Cell-Substrate Impedance Sensing (ECIS) technology. The ECIS technology is a real time cell growth assay based on a small (250 μm diameter) active gold electrode which resistance is measured continuously. When intestinal epithelial cells reach confluency on the gold electrode, resistances reach a plateau. For the wound-healing assays, confluent intestinal epithelial monolayers are subjected to a current of 40 kHz frequency, 1,400 μA amplitude, and 30-second duration. This kills the cells around the small active gold electrode, causing detachment and generating a wound that is healed by surrounding cells that have not been submitted to the current pulse. Wound healing is then assessed by continuous resistance measurements for approximately 30 h after wound. Both cell wounding and measurements of the subsequent healing process are carried out under computer control that takes online measurements each 30 s and stores the data. ECIS technology can be used to study the underlying causes for impaired mucosal healing and to test the efficacy of drugs in mucosal healing.