Dosimetric comparison of brachytherapy sources for high-dose-rate treatment of endometrial cancer: (192)Ir, (60)Co and an electronic brachytherapy source

用于子宫内膜癌高剂量率治疗的近距离放射治疗源的剂量学比较:(192)Ir、(60)Co 和电子近距离放射治疗源

阅读:1

Abstract

OBJECTIVE: To compare high-dose-rate (HDR) brachytherapy systems with (192)Ir, (60)Co and electronic brachytherapy source (EBS) for treatment of endometrial cancers. METHODS: Two additional plans were generated per patient fraction using a (60)Co source and Xoft-EBS on 10 selected patients, previously treated with a vaginal cylinder applicator using a (192)Ir source. Dose coverage of "PTV_CYLD", a 5-mm shell surrounding the cylinder, was evaluated. Doses to the following organs at risk (OARs) the rectum, bladder and sigmoid were evaluated in terms of V35% and V50%, the percentage volume receiving 35% and 50% of the prescription dose, respectively, and D2cm(3), the highest dose to a 2-cm(3) volume of an OAR. RESULTS: Xoft-EBS reduces doses to all OARs in the lower dose range, but it does not always provide better sparing of the rectum in higher dose range as does evaluation using D2cm3. V150% and V200% for PTV_CYLD was up to four times greater for Xoft-EBS plans than for plans generated with (192)Ir or (60)Co. Surface mucosal (vaginal cylinder surface) doses were also 23% higher for Xoft-EBS than for (192)Ir or (60)Co plans. CONCLUSION: Xoft-EBS is a suitable HDR source for vaginal applicator treatment with advantages of reducing radiation exposure to OARs in the lower dose range, while simultaneously increasing the vaginal mucosal dose. ADVANCES IN KNOWLEDGE: This work presents newer knowledge in dosimetric comparison between (192)Ir or (60)Co and Xoft-EBS sources for endometrial vaginal cylinder HDR planning.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。