Thermoluminescent and Monte Carlo dosimetry of IR06-103Pd brachytherapy source

IR06-103Pd近距离放射治疗源的热释光和蒙特卡罗剂量测定

阅读:1

Abstract

This work presents experimental dosimetry results for a new 103Pd brachytherapy seed, in accordance with the AAPM TG-43U1 recommendation that all new low-energy interstitial brachytherapy seeds should undergo one Monte Carlo (MC) and at least one experimental dosimetry characterization. Measurements were performed using TLD-GR200A circular chip dosimeters using standard methods employing thermoluminescent dosimeters in a Perspex phantom. The Monte Carlo N-particle (MCNP) code, version 5 was used to evaluate the dose-rate distributions around this model 103Pd source in water and Perspex phantoms. The consensus value for dose-rate constant of the IR06-103Pd source was found equal to 0.690 cGy·h(-1)·U(-1). The anisotropy function, F(r, θ), and the radial dose function, g(L)(r), of the seed were measured in Perspex phantom and calculated in both Perspex and liquid water phantom. The measured values were also found in good agreement with corresponding MC calculations.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。