Abstract
PURPOSE: Recently a number of hypothetical sources have been proposed and evaluated for use in brachytherapy. In the present study, a hypothetical (101)Rh source with mean photon energy of 121.5 keV and half-life of 3.3 years, has been evaluated as an alternative to the existing high-dose-rate (HDR) sources. Dosimetric characteristics of this source model have been determined following the recommendation of the Task Group 43 (TG-43) of the American Association of the Physicist in Medicine (AAPM), and the results are compared with the published data for (57)Co source and Flexisource (192)Ir sources with similar geometries. MATERIAL AND METHODS: MCNPX Monte Carlo code was used for simulation of the (101)Rh hypothetical HDR source design. Geometric design of this hypothetical source was considered to be similar to that of Flexisource (192)Ir source. Task group No. 43 dosimetric parameters, including air kerma strength per mCi, dose rate constant, radial dose function, and two dimensional (2D) anisotropy functions were calculated for the (101)Rh source through simulations. RESULTS: Air kerma strength per activity and dose rate constant for the hypothetical (101)Rh source were 1.09 ± 0.01 U/mCi and 1.18 ± 0.08 cGy/(h.U), respectively. At distances beyond 1.0 cm in phantom, radial dose function for the hypothetical (101)Rh source is higher than that of (192)Ir. It has also similar 2D anisotropy functions to the Flexisource (192)Ir source. CONCLUSIONS: (101)Rh is proposed as an alternative to the existing HDR sources for use in brachytherapy. This source provides medium energy photons, relatively long half-life, higher dose rate constant and radial dose function, and similar 2D anisotropy function to the Flexisource (192)Ir HDR source design. The longer half-life of the source reduces the frequency of the source exchange for the clinical environment.