Impact of different Ir-192 source models on dose calculations in high-dose-rate brachytherapy

不同Ir-192源模型对高剂量率近距离放射治疗剂量计算的影响

阅读:1

Abstract

In high-dose-rate brachytherapy, the geometry of the radioactive source is sometimes updated. Some institutions use a different source model for the dose calculation in treatment planning and treatment. The effects of this discrepancy were examined for four types of treatment plans, and ten patients were selected for each treatment plan. The impact of different source models depended on the types of treatment plan, patients, and dose index. To reduce the uncertainty and improve the reliability of the data, it would be better to use more robust metrics (D(90) and D(2cc)) for treatment planning evaluation in facilities with this problem.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。