Quantifying clinical severity of physics errors in high-dose rate prostate brachytherapy using simulations

利用模拟量化高剂量率前列腺近距离放射治疗中物理误差的临床严重程度

阅读:1

Abstract

PURPOSE: To quantitatively evaluate through automated simulations the clinical significance of potential high-dose rate (HDR) prostate brachytherapy (HDRPB) physics errors selected from our internal failure-modes and effect analysis (FMEA). METHODS AND MATERIALS: A list of failure modes was compiled and scored independently by 8 brachytherapy physicists on a one-to-ten scale for severity (S), occurrence (O), and detectability (D), with risk priority number (RPN) = SxOxD. Variability of RPNs across observers (standard deviation/average) was calculated. Six idealized HDRPB plans were generated, and error simulations were performed: single (N = 1722) and systematic (N = 126) catheter shifts (craniocaudal; -1cm:1 cm); single catheter digitization errors (tip and connector needle-tips displaced independently in random directions; 0.1 cm:0.5 cm; N = 44,318); and swaps (two catheters swapped during digitization or connection; N = 528). The deviations due to each error in prostate D90%, urethra D20%, and rectum D1cm(3) were analyzed using two thresholds: 5-20% (possible clinical impact) and >20% (potentially reportable events). RESULTS: Twenty-nine relevant failure modes were described. Overall, RPNs ranged from 6 to 108 (average ± 1 standard deviation, 46 ± 23), with responder variability ranging from 19% to 184% (average 75% ± 30%). Potentially reportable events were observed in the simulations for systematic shifts >0.4 cm for prostate and digitization errors >0.3 cm for the urethra and >0.4 cm for rectum. Possible clinical impact was observed for catheter swaps (all organs), systematic shifts >0.2 cm for prostate and >0.4 cm for rectum, and digitization errors >0.2 cm for prostate and >0.1 cm for urethra and rectum. CONCLUSIONS: A high variability in RPN scores was observed. Systematic simulations can provide insight in the severity scoring of multiple failure modes, supplementing typical FMEA approaches.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。