Source strength determination in iridium-192 and cobalt-60 brachytherapy: A European survey on the level of agreement between clinical measurements and manufacturer certificates

铱-192和钴-60近距离放射治疗中源强度的确定:一项关于临床测量与制造商证书一致性水平的欧洲调查

阅读:1

Abstract

BACKGROUND AND PURPOSE: Brachytherapy treatment outcomes depend on the accuracy of the delivered dose distribution, which is proportional to the reference air-kerma rate (RAKR). Current societal recommendations require the medical physicist to compare the measured RAKR values to the manufacturer source calibration certificate. The purpose of this work was to report agreement observed in current clinical practice in the European Union. MATERIALS AND METHODS: A European survey was performed for high- and pulsed-dose-rate (HDR and PDR) high-energy sources ((192)Ir and (60)Co), to quantify observed RAKR differences. Medical physicists at eighteen hospitals from eight European countries were contacted, providing 1,032 data points from 2001 to 2020. RESULTS: Over the survey period, 77% of the (192)Ir measurements used a well chamber instead of the older Krieger phantom method. Mean differences with the manufacturer calibration certificate were 0.01% ± 1.15% for (192)Ir and -0.1% ± 1.3% for (60)Co. Over 95% of RAKR measurements in the clinic were within 3% of the manufacturer calibration certificate. CONCLUSIONS: This study showed that the agreement level was generally better than that reflected in prior societal recommendations positing 5%. Future recommendations on high-energy HDR and PDR source calibrations in the clinic may consider tightened agreements levels.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。