Commissioning considerations for the Bravos high-dose-rate afterloader: Towards improving treatment delivery accuracy

Bravos高剂量率后装机调试注意事项:提高治疗实施精度

阅读:1

Abstract

BACKGROUND AND PURPOSE: The upgrade of major equipment can be disruptive to clinical operations and introduce risk as policy and procedures need to adapt to new technical possibilities and constraints. We describe here the transition from GammaMedPlus-iX to Bravos in a busy brachytherapy clinic, involving four afterloaders across two sites. MATERIAL AND METHODS: Our clinic employs three high-dose-rate remote afterloaders in four dedicated treatment vaults at the main site and a fourth afterloader at a regional location. Of more than 600 new HDR treatment plans performed annually, most are planned and treated intraoperatively. Most treatments are for prostate cancer, followed by GYN, intraoperative brachytherapy, GI, and other sites. Applicators used include vendor-provided applicators as well as third party applicators and in-house 3D-printed devices to provide interstitial, intracavitary, intraluminal, and surface treatments. All applicators were commissioned according to recommended guidelines. The choice of tolerances and the design of new procedures were informed by current guidelines and leveraged new HDR afterloader functionalities. A review of clinical operations in the 4 months postupgrade was conducted to evaluate the feasibility of new tolerances and the effectiveness of new procedures. RESULTS: The procedures outlined improved and standardized afterloader QA and treatment protocols with clear actionable steps for staff to follow to ensure treatments are delivered as planned. Re-commissioning of applicators yielded results similar to those previously reported by other investigators. A review of initial treatment data revealed that in one case, due to the implementation of tight tolerances, obstruction near the tip of the channel was detected and corrected prior to treatment. It confirms that the implementation of the tolerances adopted is feasible and effective in flagging treatment deviations. CONCLUSION: Enhanced procedures and QA processes were implemented successfully. We established clear actionable steps to follow by staff to ensure that treatments are delivered accurately.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。