Investigation of a source model for a new electronic brachytherapy tandem by film measurement

利用胶片测量法研究新型电子近距离放射治疗串联装置的放射源模型

阅读:1

Abstract

PURPOSE: To investigate the accuracy of a vendor-supplied source model for a new Xoft Axxent 0-degree titanium tandem by film measurement. METHODS: We measured the anisotropy factors at varying distances and angles from the tandem in water using radiochromic film (Gafchromic EBT3) and an Epson Perfection v750 desktop flatbed scanner (US Epson, Long Beach, CA). A 0-degree tandem was placed vertically in a water phantom. Four pieces of film, each at varying depths, were positioned orthogonal to the longitudinal axis of the tandem for azimuthal anisotropy measurements. Polar anisotropy measurements were taken with the film aligned parallel to the tandem. An absolute dose calibration for the film was verified with a PTW 34013 Soft X-Ray Chamber. The film measurements were analyzed using different color channels. The measured polar anisotropy for varying source positions was compared to the vendor's data. Azimuthal anisotropy was measured as a function of the radius and angle, and normalized to the mean value over all angles at the specified radius. RESULTS: The azimuthal anisotropy of the tandem and source was found to be consistent for different positions along the tandem's longitudinal axis and at varying distances from the tandem. Absolute dose using a calibrated parallel plate chamber showed agreement to within 2% of expected TPS values. The custom tandem, which has a thicker tip than the wall, was attenuating the 50 kV photons more than expected, at the angles where the photons had more wall material to traverse. This discrepancy was verified at different distances from the tandem and with different measurement techniques. As distance increased, anisotropy values had better agreement. CONCLUSIONS: We quantified the agreement between the measured and provided anisotropy factors for a new Xoft Axxent 0-degree titanium tandem. Radiochromic film response at low kV energy was also investigated. Our results showed that vendor-supplied TG-43 values were appropriate for clinical use at majority of the angles. A rigorous quality assurance method for new electronic brachytherapy sources and applicators, along with complete knowledge of all dosimetric measuring tools, should be implemented for all parts of the verification and commissioning process.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。