Real-time intraoperative computed tomography assessment of quality of permanent interstitial seed implantation for prostate cancer

实时术中计算机断层扫描评估前列腺癌永久性间质粒子植入的质量

阅读:1

Abstract

OBJECTIVES: To evaluate the use of real-time kilovoltage cone-beam computed tomography (CBCT) during prostate brachytherapy for intraoperative dosimetric assessment and correcting deficient dose regions. METHODS: A total of 20 patients were evaluated intraoperatively with a mobile CBCT unit immediately after implantation while still anesthetized. The source detector system was enclosed in a circular CT-like geometry with a bore that accommodates patients in the lithotomy position. After seed deposition, the CBCT scans were obtained. The dosimetry was evaluated and compared with the standard postimplantation CT-based assessment. In 8 patients, the deposited seeds were localized in the intraoperative CBCT frame of reference and registered to the intraoperative transrectal ultrasound images. With this information, a second intraoperative plan was generated to ascertain whether additional seeds were needed to achieve the planned prescription dose. The final dosimetry was compared with the postimplantation scan assessment. RESULTS: The mean differences between the dosimetric parameters from the intraoperative CBCT and postimplant CT scans were < .5% for percentage of volume receiving 100% of the prescription dose, minimal dose received by 90% of the prostate, and percentage of volume receiving 150% of the prescription dose. The minimal dose received by 5% (maximal dose) of the urethra differed by 8% on average and for the rectum an average difference of approximately 18% was observed. After fusion of the implanted seed coordinates from the intraoperative CBCT scans to the intraoperative transrectal ultrasound images, the dosimetric outcomes were not significantly different from the postimplantation CT dosimetric results. CONCLUSIONS: Intraoperative CT-based dosimetric evaluation of prostate permanent seed implantation before anesthesia reversal is feasible and might avert misadministration of dose delivery. The dosimetric measurements using the intraoperative CBCT scans were dependable and correlated well with the postimplant diagnostic CT findings.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。