Evaluation of a photon counting Medipix3RX cadmium zinc telluride spectral x-ray detector

对光子计数型 Medipix3RX 碲化镉锌光谱 X 射线探测器的评估

阅读:1

Abstract

We assess the performance of a cadmium zinc telluride (CZT)-based Medipix3RX energy-resolving and photon-counting x-ray detector as a candidate for spectral microcomputed tomography (micro-CT) imaging. It features an array of 128 × 128 , 110-μm2 pixels, each with four simultaneous threshold counters that utilize real-time charge summing. Each pixel's response is assessed by imaging with a range of incident x-ray intensities and detector integration times. Energy-related assessments are made by exposing the detector to the emission from an I-125 radioisotope brachytherapy seed. Long-term stability is assessed by repeating identical exposures over the course of 1 h. The high yield of properly functioning pixels (98.8%), long-term stability (linear regression of whole-chip response over 1 h of acquisitions: y =  - 0.0038x + 2284 ; standard deviation: 3.7 counts), and energy resolution [2.5 keV full-width half-maximum (FWHM) (single pixel), 3.7 keV FWHM (across the full image)] make this device suitable for spectral micro-CT.