Monte Carlo Computation of Dose-Volume Histograms in Structures at Risk of an Eye Irradiated with Heterogeneous Ruthenium-106 Plaques

利用蒙特卡罗方法计算异质钌-106斑块照射眼部风险结构中的剂量体积直方图

阅读:1

Abstract

BACKGROUND/AIMS: The aim of this work is to compare Monte Carlo simulated absorbed dose distributions obtained from (106)Ru eye plaques, whose heterogeneous emitter distribution is known, with the common homogeneous approximation. The effect of these heterogeneities on segmented structures at risk is analyzed using an anthropomorphic phantom. METHODS: The generic CCA and CCB, with a homogeneous emitter map, and the specific CCA1364 and CCB1256 (106)Ru eye plaques are modeled with the Monte Carlo code PENELOPE. To compare the effect of the heterogeneities in the segmented volumes, cumulative dose-volume histograms are calculated for different rotations of the aforementioned plaques. RESULTS: For the cornea, the CCA with the equatorial placement yields the lowest absorbed dose rate while for the CCA1364 in the same placement the absorbed dose rate is 33% higher. The CCB1256 with the hot spot oriented towards the cornea yields the maximum dose rate per unit of activity while it is 44% lower for the CCB. CONCLUSIONS: Dose calculations based on a homogeneous distribution of the emitter substance yield the lowest absorbed dose in the analyzed structures for all plaque placements. Treatment planning based on such calculations may result in an overdose of the structures at risk.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。