Constancy checks of well-type ionization chambers with external-beam radiation units

利用外束辐射装置对井型电离室进行稳定性检查

阅读:1

Abstract

Constancy checks of a well-type ionization chamber should be performed regularly as part of a quality assurance regime. The goal of this work was to test the feasibility of using a linear accelerator and an orthovoltage unit to check the constancy of a well-type chamber's response to an external radiation source. The reproducibility, linearity with dose, variation with dose-rate, and variation between energy-matched units of the well-type chamber response when exposed to 6 MV beams was examined. The robustness to errors in establishing the measurement conditions, including setting the source-to-surface distance and gantry angle, rotation of the chamber around the central axis of the beam, and the effect of changing the length of the chamber cable exposed to the field, were tested. The reproducibility and linearity with dose of the chamber response, and robustness to errors in establishing the measurement conditions for 100 kVp and 250 kVp beams from an orthovoltage unit, were also examined. The combined uncertainty, including contributions from errors in establishing the reference conditions, for well-type chamber measurements using a 6 MV beam from a linear accelerator is 1.0%. The combined uncertainties for measurements using 100 and 250 kVp beams were 1.8% and 1.5%, respectively. When focus-source distance errors were reduced to ≤ 1 mm, the combined uncertainties for the 100 and 250 kVp beams were 1.2% and 1.1%, respectively, when the dose to the chamber was confined to the linear region of the dose-response curve. The response of a well-type chamber should remain constant to within 1.2% when exposed to a constant dose from an external beam unit, if reference conditions can be reproducibly established. However, the uncertainty for establishing reference conditions for output measurements for an orthovoltage unit can be reduced, which would justify a reduction of the tolerance for constancy measurements.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。