Customisation and validation of a low-volume plasma renin activity immunoassay: Enabling of regulatory compliant determination in paediatric trials

低容量血浆肾素活性免疫测定的定制和验证:在儿科试验中实现符合法规要求的测定

阅读:6
作者:F K Suessenbach, J Tins, B B Burckhardt; LENA Consortium

Conclusion

A fit-for-purpose PRA ELISA characterised by low-volume application was successfully established, indicating non-inferiority regarding commonly applied RIAs. Reliability of the regulatory-compliant PRA assay was proven by participation in an interlaboratory ring test and its application in a paediatric population.

Methods

A PRA enzyme-linked immunosorbent assay (ELISA) was tailored for paediatric application and validated in the context of the U.S. Food and Drug Administration bioanalytical guideline. Performance verification of the assay was conducted by participation in an interlaboratory ring test, evaluation of incurred sample reanalysis and an application-orientated approach in children.

Results

A five-fold reduction of required plasma volume to 100 μL was achieved without limiting the calibration range. Between-run accuracy and precision varied no more than 5.0% and 6.3%, respectively. No substantial matrix effect was detected and the inter-run precision for parallelism was 11.1%. Stability experiments approved the freeze-thaw stability, short-term stability as well as 37 weeks of long-term stability. The assay successfully participated in the interlaboratory ring test, showing non-inferiority regarding radioimmunoassay (RIA). Moreover, PRA in plasma samples of neonates was successfully determined. Conducted incurred sample reanalysis confirmed the comparability and reliability of the assay with regard to international regulatory bioanalytical guidelines.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。