Influence of Keratoconus Severity on Detecting True Progression with Scheimpflug Imaging and Anterior Segment Optical Coherence Tomography

圆锥角膜严重程度对Scheimpflug成像和前节光学相干断层扫描检测真实进展的影响

阅读:2

Abstract

To determine consistent change over time in keratoconus disease, it is necessary to establish progression cut-off values based on intersession variability of the device used to monitor the cornea. The aim of this study was to analyze the accuracy of corneal parameters using Scheimpflug tomography and anterior segment optical coherence tomography in healthy and keratoconic eyes of varying severity to determine the cut-off values that indicate real progression. Three repeated measurements of each cornea of healthy (20 eyes) and keratoconic eyes (mild = 16, moderate = 25 and severe = 20) were recorded using Pentacam and Casia SS-1000 devices, which were repeated 2-3 weeks later. K1, K2, maximal anterior and posterior keratometry, and corneal thickness at the thinnest location (TCT) were collected. The accuracy was excellent with both devices; however, the Casia device presented better repeatability and reproducibility in all parameters in all groups compared to the Pentacam. The cut-off of the Pentacam and Casia in the mild stage were lower (K1 = 0.50 and 0.37 D; K2 = 0.51 and 0.37 D; Kmax-A = 1.24 and 0.65 D; Kmax-P = 0.38 and 0.17 D; TCT = 19.64 and 11.19 µm) than that of the severe stage (K1 = 1.09 and 0.88 D; K2 = 1.41 and 0.87 D; Kmax-A = 2.74 and 2.15 D; Kmax-P = 0.82 and 0.22 D; TCT = 28.68 and 14.83 µm). These results show that the greater the keratoconus severity, the greater the change that must occur for it to be considered real.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。