Clinical diagnostic reference levels and image quality metrics for CT in oncology patients

肿瘤患者CT临床诊断参考水平和图像质量指标

阅读:2

Abstract

BACKGROUND: In CT, the volumetric CT dose index (CTDI(vol)), dose-length product (DLP) and patient's size-specific dose estimates (SSDE) are used as diagnostic reference level (DRL) metrics. OBJECTIVES: To develop clinical local DRL values for CT chest-abdomen-pelvis (CAP) examinations using the CTDI(vol), DLP and SSDE, and to determine the image quality achieved. METHOD: In total, 201 cancer patients were included in the study. The scanning parameters, dose metrics from the CT unit and participants' body mass index (BMI) were documented. The local CT DRL values for CAP examinations were defined as the median and 75th percentiles of the dose distribution. RESULTS: The local DRL values given in terms of median CTDI(vol) ranged between 8.4 mGy and 12.7 mGy for the different types of cancers. The median DLPs ranged from 848 to 1173.4 mGy.cm for the various cancers. Generally, the radiation dose was directly proportional to the BMI and number of scan phases. Significant differences were observed between the DRLs for the various size-related parameters, number of scan phases and BMI classifications. The image quality was clinically satisfactory. CONCLUSION: No baseline data for clinical DRL values were available for this medical oncology department. The achieved DRLs were similar to published size-specific DRLs. The image quality was maintained during CT imaging. Dose optimisation and image quality assessment should be implemented to ensure optimal scanning parameters for different cancers in CT CAP examinations. CONTRIBUTION: The first size-specific local DRL values for CT CAP investigations carried out on oncology patients in South Africa have been established.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。