Abstract
OBJECTIVES: The purpose of this study was to investigate the effects of contamination of venous blood with a lipid-containing solution on parameters measured by a modern blood gas analyzer. METHODS: We collected venous blood from 17 healthcare workers (46 ± 11 years; 53 % women) into three blood gas syringes containing 0 , 5 and 10 % lipid-containing solution. Blood gas analysis was performed within 15 min from sample collection on GEM Premier 5000, while triglycerides and serum indices were assays on Roche COBAS C702. RESULTS: Triglycerides concentration increased from 1.0 ± 0.3 mmol/L in the uncontaminated blood gas syringe, to 39.4 ± 7.8 and 65.3 ± 14.4 mmol/L (both p<0.001) in syringes with 5 and 10 % final lipid contamination. The lipemic and hemolysis indices increased accordingly. Statistically significant variation was noted for all analytes except hematocrit and COHb in the syringe with 5 % lipids, while only COHb did not vary in the syringe with 10 % lipids. Significant increases were observed from 5 % lipid contamination for pO(2), SO(2) and lactate, while the values of pH, pCO(2), sodium, potassium, chloride, ionized calcium, glucose, hematocrit (10 % contamination), hemoglobin and MetHB decreased. All these changes except lactate and CoHb exceeded their relative performance specifications. CONCLUSIONS: Artifactual hyperlipidemia caused by contamination with exogenous lipids can have a clinically significant impact on blood gas analysis. Manufacturers of blood gas analyzers must be persuaded to develop new instruments equipped with serum indices.