Methods
After enzymatic hydrolysis, urinary proteins were precipitated with acetonitrile. Chromatography utilized a 10 min gradient on a Kinetex XB-C18 column with 0.1% formic acid in water and acetonitrile. Scheduled multiple reaction monitoring "survey scans" were followed by information-dependent acquisition-enhanced product ion scan experiments on an ABSciex 5500 QTRAP mass spectrometer. Analytes were identified by software-assisted library searching against reference spectra.
Results
The method was fully validated, including proof of selectivity (no exogenous or endogenous interferences were observed), assessment of matrix effects (95-122%) and recovery (53-95%), determination of limits of detection (0.5-10 ng/mL), carry-over studies (thresholds between 100 and 1000 ng/mL), and determination of autosampler stability (samples were stable for at least 3 days). Hydrolysis efficiency was thoroughly investigated for a wide range of glucuronides and for the reference standard, JWH-018 5-hydroxypentyl glucuronide.