Not all light transmission aggregation assays are created equal: qualitative differences between light transmission and 96-well plate aggregometry

并非所有透光聚集测定法都是一样的:透光聚集测定法和 96 孔板聚集测定法之间的定性差异

阅读:17
作者:Melissa V Chan, Philip D Leadbeater, Steve P Watson, Timothy D Warner

Abstract

In this short article, submitted as part of the review on platelet function testing, we illustrate the quantitative and qualitative differences between classical light transmission aggregometry (LTA) and 96-well plate aggregometry. We show that responses to platelet agonists and antagonists differ depending upon the method of aggregation testing. For example, in 96-well aggregometry, responses to collagen are strongly inhibited by P2Y12 receptor antagonists while in LTA they are much less affected. Furthermore, we explore the importance of differences in the mechanical environment upon platelet aggregation. We emphasize that LTA and 96-well aggregometry are not interchangeable assays. These two assays are best used as complementary tests to explore platelet function in depth.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。