Micro-particle image velocimetry for blood flow in thick round glass micro-channels: Channel fabrication and velocity profile characterization

厚圆形玻璃微通道中血流的微粒图像测速:通道制造和速度分布表征

阅读:8
作者:Camille Chartrand, Andy Vinh Le, Marianne Fenech

Abstract

This method describes the use of thick round borosilicate glass micro-channels for blood flow visualization using micro-particle image velocimetry (µPIV) techniques. In contrast with popular methods using squared polydimethylsiloxane channels, this method allows for visualization of blood flow in channel geometries that resemble more the natural physiology of human blood vessels. With a custom designed enclosure, the microchannels were submerged in glycerol to reduce light refraction occurring during µPIV due to the thick walls of the glass channels. A method is proposed to correct the extracted velocity profiles from the µPIV to account for out-of-focus error. The customized elements of this method include: • The use of thick circular glass micro-channels, • a custom designed mounting solution for the channels on a glass slide for flow visualization, • a MATLAB code to correct velocity profile accounting for out-of-focus error.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。