Mechanoscopy: A Novel Device and Procedure for in vivo Detection of Chronic Colitis in Mice

机械镜检查:一种用于小鼠体内慢性结肠炎检测的新型装置和方法

阅读:1

Abstract

BACKGROUND: Gut stiffening caused by fibrosis plays a critical role in the progression of inflammatory bowel disease (IBD) and colon cancer. Previous studies have characterized the biomechanical response of healthy and pathological gut, with most measurements obtained ex vivo. METHODS: Here, we developed a device and accompanying procedure for in vivo quantification of gut stiffness, termed mechanoscopy. Mechanoscopy includes a flexible balloon catheter, pressure sensor, syringe pump, and control system. The control system activates the balloon catheter and performs automated measurements of the gut stress-strain biomechanical response. RESULTS: A gut stiffness index (GSI) is identified based on the slope of the obtained stress-strain response. Using a colitis mouse model, we demonstrated that GSI positively correlates with the extent of gut fibrosis, the severity of mucosal damage, and the infiltration of immune cells. Furthermore, a critical strain value is suggested, and GSI efficiently detects pathological gut fibrotic stiffening when the strain exceeds this value. CONCLUSIONS: Based on these results, we envision that mechanoscopy and GSI will facilitate the clinical diagnosis of IBD.