Shock Lead Dislodgement Related To Its Small Hair-Pin Curve In A Pocket -A Case Of Ratchet Syndrome

电极导线移位与其在电极袋内的微小发夹弯有关——棘轮综合征病例报告

阅读:1

Abstract

There have been few reports about ratchet syndrome. We report a case of ratchet syndrome caused by small hair-pin curve of lead that triggered the lead retract itself. A 69-year-old man with a past history of inferior wall myocardial infarction, presented with progressive congestive heart failure. He underwent implantation of cardiac resynchronization therapy with an implantable cardiac defibrillator (CRTD) at our hospital. At 33 days after implantation, shock lead dislodgement was revealed. X-ray showed that the lead tip was in left subclavian vein, leaving its screw out, and a large part of the proximal portion of the lead was retracted into the pocket, while the other two leads remained in appropriate positions and the device had not rotated. An X-ray series showed that a hair-pin curve had been expanding gradually from just after implantation. In this case, relatively stiff shock lead was markedly bent and expanded the curve in the pocket, and ratchet-like movement occurred. We here report a new type of ratchet syndrome.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。