Defibrillation Threshold Testing for Right-sided Device Implants: A Review to Inform Shared Decision-making, in Association with the British Heart Rhythm Society

右侧植入式心脏装置除颤阈值测试:一项旨在为共同决策提供信息的综述(与英国心律学会合作完成)

阅读:1

Abstract

Prevention of sudden death using ICDs requires the reliable delivery of a high-energy shock to successfully terminate VF. Until more recently, the device implant procedure included conducting defibrillation threshold (DFT) testing involving VF induction and shock delivery to ensure efficacy. Large clinical trials, including SIMPLE and NORDIC ICD, have subsequently demonstrated that this is unnecessary, with a practice of omitting DFT testing having no impact on subsequent clinical outcomes. However, these studies specifically excluded patients requiring devices implanted on the right side, in whom the shock vector is significantly different and smaller studies suggest a higher DFT. In this review, the data regarding the use of DFT testing, focusing on right-sided implants, and the results of a survey of current UK practice are presented. In addition, a strategy of shared decision-making when it comes to deciding on the use of DFT testing during right-sided ICD implant procedures is proposed.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。