Optimization of image quality and patient dose in radiographs of paediatric extremities using direct digital radiography

利用直接数字放射成像技术优化儿童四肢X光片中的图像质量和患者剂量

阅读:1

Abstract

OBJECTIVE: The purpose of this study was to evaluate the effect of beam quality on the image quality (IQ) of ankle radiographs of paediatric patients in the age range of 0-1 year whilst maintaining constant effective dose (ED). METHODS: Lateral ankle radiographs of an infant foot phantom were taken at a range of tube potentials (40.0-64.5 kVp) with and without 0.1-mm copper (Cu) filtration using a Trixell Pixium 4600 detector (Trixell, Morains, France). ED to the patient was computed for the default exposure parameters using PCXMC v. 2.0 and was fixed for other beam qualities by modulating the tube current-time product. The contrast-to-noise ratio (CNR) was measured between the tibia and adjacent soft tissue. The IQ of the phantom images was assessed by three radiologists and a reporting radiographer. Four IQ criteria were defined each with a scale of 1-3, giving a maximum score of 12. Finally, a service audit of clinical images at the default and optimum beam qualities was undertaken. RESULTS: The measured CNR for the 40 kVp/no Cu image was 12.0 compared with 7.6 for the default mode (55 0.1 mm Cu). An improvement in the clinical IQ scores was also apparent at this lower beam quality. CONCLUSION: Lowering tube potential and removing filtration improved the clinical IQ of paediatric ankle radiographs in this age range. ADVANCES IN KNOWLEDGE: There are currently no UK guidelines on exposure protocols for paediatric imaging using direct digital radiography. A lower beam quality will produce better IQ with no additional dose penalty for infant extremity imaging.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。