Quality Control and Validation of Oscillometric Blood Pressure Measurements Taken During an Epidemiological Investigation

流行病学调查期间示波法血压测量的质量控制和验证

阅读:1

Abstract

This study aims to validate blood pressure (BP) values measured by an oscillometric BP monitor and seek possible calibration methods if discrepancies exist. Noninvasive BP measurement outcomes were determined using an oscillometric BP monitor (Omron HBP-1300) versus a mercury sphygmomanometer (standard device). Two percent of subjects enrolled in an epidemiological investigation were systematically sampled in this study. Intraclass correlation coefficient (ICC) was used to evaluate measurement reliability, paired t-test was used to evaluate trueness, and linear regression was used for calibration. The Association for the Advancement of Medical Instrumentation (AAMI) standards and British Hypertension Society (BHS) protocols were used for validation quality assessment. Both mercury sphygmomanometer (standard device) and oscillometric BP monitor (test device) displayed high reliability. A significant difference in systolic blood pressure (SBP) was observed between devices. SBP calibration was achieved by using an effective linear regression model (B = 0.803 and constant = 19.592, P < 0.001). The calibrated model was corroborated by verification samples (P = 0.120) and was found to pass AAMI standards and BHS protocol requirements. Calibrated SBP measurements from the Omron HBP-1300 device were valid. Use of a combination of statistical methods, such as ICC for reliability assessment as well as paired t-test for trueness evaluation can be used to validate data from the oscillometric BP monitors.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。