Improvements in carotid plaque imaging using a new eight-element phased array coil at 3T

使用新型八元件相控阵线圈在3T磁共振成像中改进颈动脉斑块成像。

阅读:1

Abstract

PURPOSE: To design and compare an eight-channel phased array (PA) coil for carotid imaging to an established four-channel PA design at 3T. MATERIALS AND METHODS: An eight-channel PA (8PA) coil was designed specifically for imaging the carotid bifurcation and compared with the existing four-channel (4PA) design using a phantom and by in vivo black-blood magnetic resonance imaging (MRI). The 8PA and 4PA were compared in terms of coverage, signal-to-noise ratio (SNR), and contrast-to-noise ratio (CNR). RESULTS: The 8PA showed up to 1.7-fold improvement in SNR at a depth of 3.5 cm and greater longitudinal coverage at a given SNR on a phantom. The 8PA showed improved vessel wall SNR for high spatial resolution (0.63 mm(2)) PD, T1, and T2 (1.7, 1.7, 1.6 times, respectively; P