Development and validation of an LC-MS/MS assay for tenofovir and tenofovir alafenamide in human plasma and cerebrospinal fluid

建立并验证用于检测人血浆和脑脊液中替诺福韦和替诺福韦艾拉酚胺的LC-MS/MS分析方法

阅读:1

Abstract

A liquid chromatography with triple quadrupole mass spectrometry method was developed and validated for the determination of tenofovir and tenofovir alafenamide concentrations in human plasma and cerebrospinal fluid. Tenofovir and tenofovir alafenamide were extracted from matrix by solid phase extraction. The dried extraction eluents were dissolved in water for LC-MS/MS analysis. Separation was achieved with a Phenomenex Synergi 4 μm Polar-RP 80A column (50 × 2 mm) with a gradient elution of 0.1% formic acid in water and acetonitrile. The total run time was 5 min. Detection of analytes was achieved using electrospray ionization (positive mode) and triple quadrupole selected reaction monitoring. Standard curve concentrations ranged from 0.5 to 500 ng/mL for the plasma assay and 0.1-50 ng/mL for the cerebrospinal fluid assay. The intra- and inter-day accuracy and precision were less than 12% in low, medium, and high quality control samples for both matrices. The validated methods were applied to the analysis of plasma and cerebrospinal fluid samples of a patient undergoing tenofovir therapy which involved the switch from Stribild(®) (elvitegravir 150 mg/cobicistat 150 mg/emtricitabine 200 mg/tenofovir disoproxil fumarate 300 mg) to Genvoya(®) (elvitegravir 150 mg/cobicistat 150 mg/emtricitabine 200 mg/tenofovir alafenamide 10 mg).

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。