Application of UPLC-MS/MS for separation and quantification of 3α-Hydroxy Tibolone and comparative bioavailability of two Tibolone formulations in healthy volunteers

应用超高效液相色谱-串联质谱法分离和定量3α-羟基替勃龙,并比较两种替勃龙制剂在健康志愿者中的生物利用度

阅读:1

Abstract

A novel, fast, sensitive and robust method based on ultra-performance liquid chromatography coupled to atmospheric pressure electrospray ionization tandem mass spectrometry (UPLC-ESI-MS/MS) has been developed to separate two Tibolone stereoisomers i.e., 3α-Hydroxy Tibolone and 3β-Hydroxy Tibolone and to quantify 3α-Hydroxy Tibolone using p-toulenesulfonyl isocyanate (PTSI) as a derivatizing reagent in human plasma. 3α-Hydroxy Tibolone-(13)CD(3) was used as an internal standard (IS). The analyte and IS were extracted from human plasma by liquid-liquid extraction using ethyl acetate. Extracted samples were analyzed by UPLC-ESI-MS/MS. Chromatography was performed using binary gradient on UPLC analytical column. A linear calibration curve over the range of 0.100-35.000 ng/mL was obtained and lower limit of quantification (LLOQ) was 0.100 ng/mL demonstrating acceptable accuracy and precision. This method was successfully applied to a pharmacokinetic study in order to compare a test Tibolone 2.5 mg formulation vs. a reference 2.5 mg Tibolone tablet formulation in 50 post-menopausal/surgical menopause female human volunteers under fasting conditions. It is concluded that test formulation of Tibolone is bioequivalent to reference formulation of Tibolone.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。