Quantification for non-targeted LC/MS screening without standard substances

无需标准物质的非靶向液相色谱/质谱筛选定量分析

阅读:1

Abstract

Non-targeted and suspect analyses with liquid chromatography/electrospray/high-resolution mass spectrometry (LC/ESI/HRMS) are gaining importance as they enable identification of hundreds or even thousands of compounds in a single sample. Here, we present an approach to address the challenge to quantify compounds identified from LC/HRMS data without authentic standards. The approach uses random forest regression to predict the response of the compounds in ESI/HRMS with a mean error of 2.2 and 2.0 times for ESI positive and negative mode, respectively. We observe that the predicted responses can be transferred between different instruments via a regression approach. Furthermore, we applied the predicted responses to estimate the concentration of the compounds without the standard substances. The approach was validated by quantifying pesticides and mycotoxins in six different cereal samples. For applicability, the accuracy of the concentration prediction needs to be compatible with the effect (e.g. toxicology) predictions. We achieved the average quantification error of 5.4 times, which is well compatible with the accuracy of the toxicology predictions.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。