Analysis of Induced Field in the Brain Tissue by Transcranial Magnetic Stimulation Using Halo-V Assembly Coil

利用Halo-V组装线圈进行经颅磁刺激分析脑组织中的感应场

阅读:1

Abstract

As a noninvasive neuromodulation technique, transcranial magnetic stimulation (TMS) has already exhibited a great impact in clinical application and scientific research. This study presents a finite element method-based simulation of the Halo-V assembly (HVA) coil placed on the five-shell spherical human head model to examine the distributions of induced electric and magnetic fields. The performance of the designed HVA coil is evaluated by comparing the simulation results with the commercially available Halo-FO8 (HFA) assembly coil and standard single coils including the Halo and V coils. The simulation results indicate that the HVA coil shows an improved focality in terms of electric field distribution than the other single and assembly stimulation coils. Additionally, the effects of a magnetic shield plate and magnetic core on the designed HVA coil are investigated. Results indicate that the magnetic shield plate and magnetic core are proficient in further improving the stimulation focality. Therefore, the HVA TMS coil results in a safe and effective stimulation with enhanced focality of the target region as compared to the existing assembly coil.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。