Lateral flow-based antibody testing for Chlamydia trachomatis

沙眼衣原体横向流动抗体检测

阅读:5
作者:Sarah Gwyn, Alexandria Mitchell, Deborah Dean, Harran Mkocha, Sukwan Handali, Diana L Martin

Abstract

We describe here a lateral flow-based assay (LFA) for the detection of antibodies against immunodominant antigen Pgp3 from Chlamydia trachomatis, the causative agent of urogenital chlamydia infection and ocular trachoma. Optimal signal detection was achieved when the gold-conjugate and test line contained Pgp3, creating a dual sandwich capture assay. The LFA yielded positive signals with serum and whole blood but not with eluted dried blood spots. For serum, the agreement of the LFA with the non-reference multiplex assay was 96%, the specificity using nonendemic pediatric sera was 100%, and the inter-rater agreement was κ=0.961. For whole blood, the agreement of LFA with multiplex was 81.5%, the specificity was 100%, and the inter-rater agreement was κ=0.940. The LFA was tested in a field environment and yielded similar results to those from laboratory-based testing. These data show the successful development of a lateral flow assay for detection of antibodies against Pgp3 with reliable use in field settings, which would make antibody-based testing for trachoma surveillance highly practical, especially after cessation of trachoma elimination programs.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。