The impact of differences in plasma glucose between glucose oxidase and hexokinase methods on estimated gestational diabetes mellitus prevalence

葡萄糖氧化酶法和己糖激酶法测定血浆葡萄糖差异对妊娠期糖尿病患病率估计的影响

阅读:1

Abstract

We evaluated the extent of measurement discordance between glucose oxidase and hexokinase laboratory methods and the effect of this on estimated gestational diabetes mellitus (GDM) prevalence in a routine clinical setting. 592 consecutive urban African women were screened for GDM. Paired venous specimens were submitted to two independent calibrated laboratories that used either method to measure plasma glucose concentrations. World Health Organisation diagnostic criteria were applied. GDM prevalence determined by the glucose oxidase and hexokinase methods was 6.9% and 5.1% respectively. The overall GDM prevalence was 9%. Only 34% of GDM positive diagnoses were common to both laboratory methods. Bland Altman plots identified a bias of 0.2 mmol/l between laboratory methods. Plasma glucose concentrations measured by the glucose oxidase method were more platykurtic in distribution. Low diagnostic agreement between laboratory methods was further indicated by a Cohen's kappa of 0.48 (p < 0.001). Reports of GDM prevalence using either the glucose oxidase or hexokinase laboratory methods may not be truly interchangeable or directly comparable.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。