Investigation of clinically acceptable agreement between two methods of automatic measurement of limb occlusion pressure: a randomised trial

比较两种肢体阻断压自动测量方法在临床上可接受的一致性:一项随机试验

阅读:1

Abstract

BACKGROUND: Development of automatic, pneumatic tourniquet technology and use of personalised tourniquet pressures has improved the safety and accuracy of surgical tourniquet systems. Personalisation of tourniquet pressure requires accurate measurement of limb occlusion pressure (LOP), which can be measured automatically through two different methods. The 'embedded LOP' method measures LOP using a dual-purpose tourniquet cuff acting as both patient sensor and pneumatic effector. The 'distal LOP' method measures LOP using a distal sensor applied to the patient's finger or toe of the operating limb, using photoplethysmography to detect volumetric changes in peripheral blood circulation. The distal LOP method has been used clinically for many years; the embedded LOP method was developed recently with several advantages over the distal LOP method. While both methods have clinically acceptable accuracy in comparison to LOP measured using the manual Doppler ultrasound method, these two automatic methods have not been directly compared. The purpose of this study is to investigate if the embedded and distal methods of LOP measurement have clinically acceptable agreement. The differences in pairs of LOP measurement in the upper and lower limbs of 81 healthy individuals were compared using modified Bland and Altman analysis. In surgery, it is common for cuff pressure to deviate from the pressure setpoint due to limb manipulation. Surgical tourniquet systems utilise a ± 15 mmHg pressure alarm window, whereby if the cuff pressure deviates from the pressure setpoint by > 15 mmHg, an audiovisual alarm is triggered. Therefore, if the difference (bias) ± SE, 95% CI of the bias and SD of differences ± SE in LOP measurement between the embedded and distal methods were all within ±15 mmHg, this would demonstrate that the two methods have clinically acceptable agreement. RESULTS: LOP measurement using the embedded LOP method was - 0.81 ± 0.75 mmHg (bias ± standard error) lower than the distal LOP method. The 95% confidence interval of the bias was - 2.29 to 0.66 mmHg. The standard deviation of the differences ± standard error was 10.35 ± 0.49 mmHg. These results show that the embedded and distal methods of LOP measurement demonstrate clinically acceptable agreement. CONCLUSIONS: The findings of this study demonstrate clinically acceptable agreement between the embedded and distal methods of LOP measurement. The findings support the use of the embedded LOP method of automatic LOP measurement using dual-purpose tourniquet cuffs to enable accurate, effective and simple prescription of personalised tourniquet cuff pressures in a clinical setting.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。