Metrological Comparison of Indirect Calibration Methods for Nanoindentation: A Bootstrap-Based Approach

纳米压痕间接校准方法的计量学比较:一种基于自举法的方法

阅读:1

Abstract

Area shape function and frame compliance are the most critical parameters in nanoindentation, as they control measurement accuracy and represent the largest contributions to measurement uncertainty. Despite the availability of direct calibration methods, indirect calibrations are the most practical and fast. Thus, the indirect calibration methods proposed in ISO 14577-2 are most typically applied in academic and industrial research, as well as in quality controls. Previous research has highlighted some criticalities, but a holistic metrological framework was missing. This work aims to compare the performances of indirect calibration methods for area shape function and frame compliance in the nano-range, considering different alternatives suggested in the standard and most recent literature. The comparison will be based on uncertainty estimation using bootstrap estimation, which will innovatively highlight and introduce the effect of the nanoindentation dataset in the uncertainty estimation. The results show that the optimization of accuracy and uncertainty in mechanical characterization is achieved by indenting pairs of certified reference materials, resulting in a more robust approach to calibration experimental conditions than methods that require a single sample to be indented.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。