Enhanced Sensing Behavior of Three-Dimensional Microfluidic Paper-Based Analytical Devices (3D-μPADs) with Evaporation-Free Enclosed Channels for Point-of-Care Testing

用于即时检测的具有无蒸发封闭通道的三维微流控纸基分析装置(3D-μPADs)的增强型传感性能

阅读:1

Abstract

Despite the potential in fabrication of microfluidic paper-based analytical devices (μPADs) for point-of-care testing (POCT) kits, the development of simple, accurate, and rapid devices with higher sensitivity remains challenging. Here, we report a novel method for 3D-μPAD fabrication with enclosed channels using vat photopolymerization to avoid fluid evaporation. In detail, height of the enclosed channels was adjusted from 0.3 to 0.17 mm by varying the UV exposure time from 1 to 4 s for the top barrier, whereas the exposure time for the bottom and side barriers was fixed. As a result, sample flow in the enclosed channels of 3D-μPADs showed lesser wicking speed with very scant evaporation compared to that in the hemi channels in the 3D-μPADs. The stoppage of evaporation in the enclosed channels significantly improved the gray intensity and uniformity in the detection zone of the 3D-μPADs, resulting in as low as 0.3 mM glucose detection. Thus 3D-μPADs with enclosed channels showed enhanced sensitivity compared to the 3D-μPADs with hemi channels when dealing with a small volume sample. Our work provides a new insight into 3D-μPAD design with enclosed channels, which redefines the methodology in 3D printing.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。