Reliability and validity of an app-assisted tissue compliance meter in measuring tissue stiffness on a phantom model

应用辅助组织顺应性仪在测量体模组织硬度方面的可靠性和有效性

阅读:1

Abstract

BACKGROUND: Most methods for soft tissue stiffness assessment require high financial resources, significant technical effort, or extensive therapist training. The PACT Sense device was developed to be used in a wide range of applications and user groups. However, to date, there are no data on its validity and reliability. The aim of this study was to investigate the validity and reliability of the PACT device. METHODS: A polyurethane phantom tissue model (PTM) mimicking the mechanical properties of the fascia profunda and the erector spinae muscle was used. Stiffness measurements with PACT were conducted by two independent investigators. For construct validity, correlations were calculated between the known stiffness of the PTM and values obtained with PACT. For concurrent validity, we determined the association between the PACT values and additional measurements with the established MyotonPRO device. To estimate interrater and intrarater (two measurements with an interval of 7 days) reliability, we used the intraclass correlation coefficient (ICC). RESULTS: Correlation analysis (PTM/PACT) revealed very high concurrent validity (r = 0.99; p < 0.001), construct validity (PACT/MyotonPRO) was 0.87, p < 0.001. Both, interrater reliability (ICC = 0.85; p = 0.036) and intrarater reliability were good (ICC = 0.89; p < 0.001). CONCLUSIONS: The PACT provides valid and reliable stiffness measurements in tissue phantoms. Further studies in humans are needed to confirm its physiometric properties under in vivo conditions.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。