Clinical method evaluation of hemoglobin S and C identification by top-down selected reaction monitoring and electron transfer dissociation

通过自上而下选择反应监测和电子转移解离法对血红蛋白S和C进行鉴定的临床方法评价

阅读:1

Abstract

BACKGROUND: Biological diagnosis of hemoglobin disorders is a complex process relying on the combination of several analytical techniques to identify Hb variants in a particular sample. Currently, hematology laboratories usually use high-performance liquid chromatography (HPLC), capillary electrophoresis and gel-based methods to characterize Hb variants. Co-elution and co-migration may represent major issues for precise identification of Hb variants, even for the most common ones such as Hb S and C. METHODS: We adapted a top-down selected reaction monitoring (SRM) electron transfer dissociation (ETD) mass spectrometry (MS) method to fit with a clinical laboratory environment. An automated analytical process with semi-automated data analysis compatible with a clinical practice was developed. A comparative study between a reference HPLC method and the MS assay was performed on 152 patient samples. RESULTS: The developed workflow allowed to identify with high specificity and selectivity the most common Hb variants (Hb S and Hb C). Concordance of the MS-based approach with HPLC was 71/71 (100%) for Hb S and 11/11 (100%) for Hb C. CONCLUSIONS: This top-down SRM ETD method can be used in a clinical environment to detect Hb S and Hb C.