A comparison of the analysis of 3 types of body fluids using the XN-350 hematology analyzer versus light microscopy assessment

比较使用 XN-350 血液分析仪和光学显微镜评估方法分析 3 种体液的结果

阅读:1

Abstract

We evaluated the capacity of the XN-350 instrument to analyze 3 different types of body fluid samples under "body fluid mode."The performance of XN-350 was evaluated in terms of precision, carryover, limit of blank, limit of detection, limit of quantification, and linearity. Cell enumeration and differential data produced by the XN-350 were compared to manual chamber counting results in 63 cerebrospinal fluid (CSF), 51 ascitic fluid, and 51 pleural fluid (PF) samples. Comparisons between XN-350 versus Cytospin data were also performed in PF samples.The precision, carry-over, limit of blank, and linearity of the XN-350 were acceptable. The limits of detection for white blood cells (WBCs) and red blood cells were 1.0/μL, and 1,000.0/μL, respectively; the corresponding limits of quantitation (LOQs) were 5.0/μL and 2,000.0/μL, respectively. The XN-350's cell enumeration and differential counting correlated well with those of manual chamber counting for all 3 sample types (except for differential counting in CSF samples), particularly parameters involving monocytes (r = 0.33) and mononuclear cells (MO- body fluid [BF]; r = 0.26), as well as total cell (TC-BF) enumeration (r = 0.50) and WBC-BF (r = 0.50) in PF samples. The MO-BF in CSF samples differed significantly from manual chamber counting results, but neither TC-BF nor WBC-BF in PF samples did. The XN-350 also showed good correlations with Cytospin analyses for differential counting of neutrophils, lymphocytes, and monocytes in PF samples. The differential counting of eosinophils via the XN-350 and Cytospin were not significantly correlated, but the difference between them was not significant.The XN-350 is an acceptable alternative to manual fluid analysis. Samples with low cellularity around the LOQ should be checked manually. Moreover, manual differential counting should be performed on CSF samples, particularity those with low cell numbers.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。