Solving selectivity issues in LBAs: case study using Gyrolab to quantify CB307, a bispecific Humabody in human serum

解决 LBA 中的选择性问题:使用 Gyrolab 量化人血清中的双特异性抗体 CB307 的案例研究

阅读:13
作者:Thomas Wilford, Phillip D Bartlett, Anna Schlag, Lukas Jasaitis, Hardev Pandha, Andrew J Pierce, Richard Hughes

Aim

Endogenous interferents can cause nonselectivity in ligand binding pharmacokinetic assays, leading to inaccurate quantification of drug concentrations. We describe the development of a Gyrolab immunoassay to quantify a new modality, CB307 and discuss strategies implemented to overcome matrix effects and achieve selectivity at the desired sensitivity.

Conclusion

The strategies described resulted in a selective method for CB307 in disease state matrix that met bioanalytical method validation (BMV) guidance and is currently used to support clinical pharmacokinetic sample analysis in the first-in-human POTENTIA clinical study (NCT04839991) as a secondary clinical end point.

Results

Matrix effects were mitigated using strategies including increasing minimum required dilution (MRD) and lower limit of quantification, optimization of antibody orientation, assay buffer and solid phase.