Quantification of Iodine Concentration Using Single-Source Dual-Energy Computed Tomography in a Calf Liver

利用单源双能计算机断层扫描技术定量分析小牛肝脏中的碘浓度

阅读:1

Abstract

OBJECTIVE: To evaluate the accuracy of single-source dual-energy computed tomography (ssDECT) in iodine quantification using various segmentation methods in an ex vivo model. METHODS: Ten sausages, injected with variable quantities of iodinated contrast, were inserted into 2 livers and scanned with ssDECT. Material density iodine images were reconstructed. Three radiologists segmented each sausage. Iodine concentration, volume, and absolute quantity were measured. Agreement between the measured and injected iodine was assessed with the concordance correlation coefficient (CCC). Intrareader agreement was assessed using the intraclass correlation coefficient (ICC). RESULTS: Air bubbles were observed in sausage (IX). Sausage (X) was within the same view as hyper-attenuating markers used for localization. With IX and X excluded, CCC and ICC were greater than 0.98 and greater than 0.88. When included, CCC and ICC were greater than 0.94 and greater than 0.79. CONCLUSIONS: Iodine quantification was reproducible and precise. However, accuracy reduced in sausages consisting of air filled cavities and within the same view as hyperattenuating markers.