Quantitative evaluation of breast density using a dual-energy technique on a digital breast tomosynthesis system

利用数字乳腺断层合成系统上的双能技术对乳腺密度进行定量评估

阅读:1

Abstract

PURPOSE: Although breast density is considered a strong risk factor of breast cancer, its quantitative assessment is difficult. To investigate a quantitative method of measuring breast density using dual-energy mammographic imaging with central digital breast tomosynthesis in physically uniform and nonuniform phantoms. MATERIAL AND METHODS: The dual-energy imaging unit used a tungsten anode and silver filter with 30 kVp for high-energy images and 20 kVp for low-energy images. Uniform glandular-equivalent phantoms were used to calibrate a dual-energy based decomposition algorithm. The first study used uniform breast phantoms which ranged in thicknesses from 20 to 70 mm, in 10-mm increments, and which provided 30%, 50%, and 70% of breast density. The second study used uniform phantoms ranging from 10% to 90% of breast density. The third study used non-uniform phantoms (at an average density of 50%) with a thickness which ranged from 20 to 90 mm, in 10-mm increments. RESULTS: The root mean square error of breast density measurements was 2.64-3.34% for the uniform, variable thickness phantoms, 4.17% for the uniform, variable density phantoms, and 4.49% for the nonuniform, variable thickness phantoms. CONCLUSION: The dual-energy technique could be used to measure breast density with a margin of error of < 10% using digital breast tomosynthesis.