Theoretical limits on brain cooling by external head cooling devices

外部头部冷却装置对大脑冷却的理论限制

阅读:2

Abstract

Numerous experimental studies have demonstrated that mild hypothermia is a rather promising therapy for acute brain injury in neonates. Because measurement of the resultant cooling of human brain in vivo is beyond current technology, an understanding of physical factors limiting the possible brain cooling would be a substantial achievement. Herein brain cooling by external head cooling devices is studied within the framework of an analytical model of temperature distribution in the brain. Theoretical limits on brain hypothermia induced by such devices are established. Analytical expressions are obtained that allow evaluation of changes in brain temperature under the influence of measurable input parameters. We show that a mild hypothermia can be successfully induced in neonates only if two necessary conditions are fulfilled: sufficiently low cerebral blood flow and sufficiently high value of the heat transfer coefficient describing the heat exchange between the head surface and a cooling device.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。