Abstract
In the past we have reported significant cognitive deficits in mice receiving 5-fluorouracil in combination with low-dose methotrexate. To explain such interactions, a pharmacokinetic study was designed. A sensitive bio-analytical method was therefore developed and validated for 5-fluorouracil and methotrexate in mouse plasma, brain and urine with liquid chromatography coupled to a single quadrupole mass spectrometer. Chromatographic separation was accomplished by Agilent® Zorbax® SB-C18 column, with isocratic elution (5 mM ammonium acetate and methanol, 70:30, %v/v) at a flow rate of 300 μL/min. The limit of quantitation for both drugs was 15.6 ng/mL (plasma and brain) and 78.1 ng/mL (urine), with interday and intraday precision and accuracy ≤15% and a total run time of 6 min. This bio-analytical method was used for the pharmacokinetic characterization of 5-fluorouracil and methotrexate in mouse plasma, brain and urine over a period of 24 h. This method allowed characterization of the brain concentrations of 5-fluorouracil over a period of 24 h.