Doses delivered by portal imaging quality assurance in routine practice of adjuvant breast radiotherapy worth to by monitored and compensated in some cases

在辅助乳腺放射治疗的常规实践中,通过门控成像质量保证所输送的剂量值得监测,在某些情况下还需要进行补偿。

阅读:1

Abstract

BACKGROUND: Imaging, in radiotherapy, has become a routine tool for repositioning of the target volume at each session. The repositioning precision, currently infracentimetric, evolves along with the irradiation techniques. This retrospective study aimed to identify practices and doses resulting from the use of high energy planar imaging (portal imaging) in daily practice. METHODS: A retrospective survey of portal images (PIs) was carried out over 10 years for 2,403 patients and for three linacs (1 Elekta SLi, 2 Varian Clinac) for postoperative mammary irradiations. Images were taken using a standardized number of monitor units (MU) for all patients. Due to the variable sensitivities of the detectors and the possibility of adjustment of the detector-patient distance, the number of MU were 3; 2 and 1 respectively, for Elekta SLi(®), Clinac 600(®) and Clinac 2100(®). Then, a representative cumulated dose was calculated in simplified reference conditions (5 cm depth, beam of 10 cm × 10 cm, 6 MV), considering the total number of images taken during the whole treatment course. The consistency between the representative doses and the actual absorbed doses received by the patients was verified by simulating a series of typical cases with the treatment plan dose calculation system. RESULTS: The delivered doses differ significantly between the three linacs. The mean representative dose values by complete treatment were 0.695; 0.241 and 0.216 Gy, respectively, for SLi, Clinac 600 and Clinac 2100. However, 15 patients were exposed to a dose >2 Gy with a maximum dose of 5.05 Gy. The simulated doses were very similar to the representative doses. CONCLUSIONS: A significant dose delivery was highlighted by this study. These representative doses are presently communicated weekly to the radiation oncologist for the radiation protection of their patients. Moreover, they should be taken into account in a possible study of long-term stochastic risks.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。