Dose uncertainties associated with a set density override of unknown hip prosthetic composition

与未知髋关节假体成分的设定密度覆盖相关的剂量不确定性

阅读:1

Abstract

The dosimetric uncertainties associated with radiotherapy through hip prostheses while overriding the implant to a set density within the TPS has not yet been reported. In this study, the uncertainty in dose within a PTV resulting from this planning choice was investigated. A set of metallic hip prosthetics (stainless steel, titanium, and two different Co-Cr-Mo alloys) were CT scanned in a water bath. Within the TPS, the prosthetic pieces were overridden to densities between 3 and 10 g/cm(3) and irradiated on a linear accelerator. Measured dose maps were compared to the TPS to determine which density was most appropriate to override each metal. This was shown to be in disagreement with the reported literature values of density which was attributed to the TPS dose calculation algorithm and total mass attenuation coefficient differences in water and metal. The dose difference was then calculated for a set density override of 6 g/cm(3) in the TPS and used to estimate the dose uncertainty beyond the prosthesis. For beams passing through an implant, the dosimetric uncertainty in regions of the PTV may be as high as 10% if the implant composition remains unknown and a set density override is used. These results highlight limitations of such assumptions and the need for careful consideration by radiation oncologist, therapist, and physics staff.