Acoustic Immunosensing of Exosomes Using a Quartz Crystal Microbalance with Dissipation Monitoring

使用带有耗散监测的石英晶体微天平对外泌体进行声学免疫传感

阅读:5
作者:Jugal Suthar, Edward S Parsons, Bart W Hoogenboom, Gareth R Williams, Stefan Guldin

Abstract

Exosomes are endocytic lipid-membrane bound bodies with the potential to be used as biomarkers in cancer and neurodegenerative disease. The limitations and scarcity of current exosome characterization approaches have led to a growing demand for translational techniques, capable of determining their molecular composition and physical properties in physiological fluids. Here, we investigate label-free immunosensing, using a quartz crystal microbalance with dissipation monitoring (QCM-D), to detect exosomes by exploiting their surface protein profile. Exosomes expressing the transmembrane protein CD63 were isolated by size-exclusion chromatography from cell culture media. QCM-D sensors functionalized with anti-CD63 antibodies formed a direct immunoassay toward CD63-positive exosomes in 75% v/v serum, exhibiting a limit-of-detection of 2.9 × 108 and 1.4 × 108 exosome sized particles (ESPs)/mL for frequency and dissipation response, respectively, i.e., clinically relevant concentrations. Our proof-of-concept findings support the adoption of dual-mode acoustic analysis of exosomes, leveraging both frequency and dissipation monitoring for use in bioanalytical characterization.